LOCATOR¿ Implant Attachment System. The LOCATOR¿ Implant Attachment System is a universal hing...

FDA Device Recall #Z-0935-2014 — Class II — January 16, 2014

Recall Summary

Recall Number Z-0935-2014
Classification Class II — Moderate risk
Date Initiated January 16, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Zest Anchors LLC
Location Escondido, CA
Product Type Devices
Quantity 3,130

Product Description

LOCATOR¿ Implant Attachment System. The LOCATOR¿ Implant Attachment System is a universal hinge, resilient attachment for endosseous implants. The LOCATOR¿ Implant Attachment System is designed for use with overdentures or partial dentures, retained in whole or in part, by endosseous implants in the mandible or maxilla. FDA 510(k): K994257, K072878

Reason for Recall

Zest Anchors, LLC is voluntarily recalling specific lots of LOCATOR¿ Implant Attachment Systems because it is currently provided with a label that indicates that the final torque applied to the abutment should be 30 N-cm. It was recently noted by Zest that in order to provide an optimal performance by increasing the factor of safety between the failure load of the abutment screw and the recommend

Distribution Pattern

Distributed in Ohio and Switzerland.

Lot / Code Information

Part Numbers: 8875-SPI, 8875-SPI-TMS, 8876-SPI, 8876-SPI-TMS, 08877-SPI, 08877-SPI-TMS, 08878-SPI, 08878-SPI-TMS, 08879-SPI, 08880-SPI; Lot Numbers: 11052, 15306, 20665, 21351, 12025,12259, 22440, I0K7X, I0K7Y, 11963,15307,17676,18934, 25294, 25567, I0JQ7, 12258,12941,15000, 19735, 20666, I0K7Z, I0KY7, 12703, 14568, 19832, 22170, I0LCG, 11093, 13364, 14151, 15624, 20043, 25026, 13530, I0LE4, 10055, 14152, 15001, 25027, 10892, 10893. Since the LOCATOR¿ Implant Attachment System (including the restorative components) is not sold by the Zest Anchors as a sterilized product, it does not have a designated shelf-life.

Other Recalls from Zest Anchors LLC

Recall # Classification Product Date
Z-2425-2020 Class II LOCATOR OVERDENTURE IMPLANT SYSTEM STERILE R ... Nov 22, 2019

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.