Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pres...
FDA Device Recall #Z-3220-2026 - Class II, August 21, 2026
Recall Summary
| Recall Number | Z-3220-2026 |
| Classification | Class II - Moderate risk |
| Date Initiated | August 21, 2026 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Midmark Corporation |
| Location | Versailles, OH |
| Product Type | Devices |
| Quantity | 33 units |
Product Description
Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients, Model Number: 1-200-0360 with Kit Number: 4-000-0724
Reason for Recall
A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used.
Distribution Pattern
US Nationwide distribution.
Lot / Code Information
Model No: 1-200-0360; UDI-DI: 00841709105838; Kit No: 4-000-0724; UDI-DI: 00841709105913; Kit Serial Number: V2788703 V2788046 V2788016 V2788712 V2789302 V2789846 V2789830 V2789837 V2789808 V2789303 V2789304 V2811103 V2811112 V2862288 V2862275 V2869139 V2866577 V2867451 V2869599 V2868801 V2869154 V2868545 V2866984 V2869980 V2869969 V2868130 V2867433 V2868182 V2924525 V2924566 V2924559 V2924539 V2927525; Zone Serial Number: V2788322 V2787552 V2787555 V2788321 V2788905 V2789498 V2787554 V2789488 V2789499 V2788922 V2787553 V2807891 V2807893 V2837260 V2859050 V2868888 V2862992 V2866942 V2868893 V2868239 V2868840 V2868222 V2866940 V2868892 V2869397 V2867491 V2866963 V2868051 V2923639 V2922999 V2922998 V2923640 V2927201
Other Recalls from Midmark Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-3218-2026 | Class II | Midmark IQvitals Zone, automatically and noninv... | Aug 21, 2026 |
| Z-3219-2026 | Class II | Midmark IQvitals Zone, automatically and noninv... | Aug 21, 2026 |
| Z-3217-2026 | Class II | Midmark IQvitals Zone, automatically and noninv... | Aug 21, 2026 |
| Z-3013-2026 | Class II | Midmark Digital ECG HD+12, HD+15, Model/Catalog... | Jul 20, 2026 |
| Z-3012-2026 | Class II | Midmark IQecg Digital ECG, Model/Catalog Number... | Jul 20, 2026 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.