VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.

FDA Recall #Z-0327-2026 — Class II — October 16, 2025

Recall #Z-0327-2026 Date: October 16, 2025 Classification: Class II Status: Ongoing

Product Description

VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.

Reason for Recall

Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.

Recalling Firm

Canon Medical System, USA, INC. — Tustin, CA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

2 units

Distribution

US distribution to Ohio and Nevada

Code Information

CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. Serial Numbers: B7A2462002, B7A2522014.

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated