Philips DigitalDiagnost X-Ray System with Eleva Software 2.1.3 Model numbers 712020, 712022, and...
FDA Device Recall #Z-0884-2013 — Class II — October 25, 2012
Recall Summary
| Recall Number | Z-0884-2013 |
| Classification | Class II — Moderate risk |
| Date Initiated | October 25, 2012 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Philips Healthcare Inc. |
| Location | Andover, MA |
| Product Type | Devices |
| Quantity | 551 |
Product Description
Philips DigitalDiagnost X-Ray System with Eleva Software 2.1.3 Model numbers 712020, 712022, and 712082 Product Usage: Digital radiographic system
Reason for Recall
When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image.
Distribution Pattern
Worldwide Distribution - USA Nationwide and the Foreign countries of: Austria, Canada, Belgium, China,Denmark, France, germany, Italy, Netherlands,New Zealand, Norway, Philippines, Portugual, SOuth Korea, Spain, Switzerland, Thailand, and UK.
Lot / Code Information
Systems with Eleva SW version 2.1.3
Other Recalls from Philips Healthcare Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2059-2013 | Class II | Philips Easy Upgrade DR, Code No: 712086 Thi... | Aug 8, 2013 |
| Z-1780-2013 | Class II | Radiohead Access Point 1.4 GHz Radiohead Access... | Jul 8, 2013 |
| Z-1797-2013 | Class II | Philips HeartStart MRx Monitor/Defibrillator, ... | Jun 19, 2013 |
| Z-1964-2013 | Class II | Philips Healthcare MobileDiagnostwDR system. M... | Jun 11, 2013 |
| Z-1189-2013 | Class II | HeartStart MRx Monitor/Defribillator; Product C... | Apr 8, 2013 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.