SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# F...

FDA Recall #Z-1128-2013 — Class II — December 13, 2012

Recall #Z-1128-2013 Date: December 13, 2012 Classification: Class II Status: Terminated

Product Description

SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale is designed for Joerns Healthcare for use in portable patient weighing systems. The scale is designed to be connected between the patient hoist mechanism and patient sling apparatus. The scale is not intended to be used for patient transport.

Reason for Recall

SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale. The failures resulted in patient falls.

Recalling Firm

Sr Instruments Inc — Tonawanda, NY

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

806 patient scales

Distribution

USA Nationwide Distribution in the state of Texas.

Code Information

Serial numbers: 1005 thru 1780, 1785 thru 1809, 1866 thru 1868 and 1870 thru 1875.

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated