Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX ...
FDA Device Recall #Z-2612-2026 - Class II, May 28, 2026
Recall Summary
| Recall Number | Z-2612-2026 |
| Classification | Class II - Moderate risk |
| Date Initiated | May 28, 2026 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Medline Industries, LP |
| Location | Northfield, IL |
| Product Type | Devices |
| Quantity | 9,272 kits |
Product Description
Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530A
Reason for Recall
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Distribution Pattern
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Lot / Code Information
UDI-DI 10195327033620 Lots 21KBN891 21KBO410 22ABA676 22ABY937 22BBN909 22BBT447 23ABO303
Other Recalls from Medline Industries, LP
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2696-2026 | Class II | Medline Convenience Kits: 1) OPEN HEART PEDS P... | May 28, 2026 |
| Z-2697-2026 | Class II | Medline Convenience Kits: 1) OPEN HEART PEDS P... | May 28, 2026 |
| Z-2695-2026 | Class II | Medline Convenience Kits: 1) ANESTHESIA KIT PE... | May 28, 2026 |
| Z-2610-2026 | Class II | Medline Convenience kits containing recalled Sw... | May 28, 2026 |
| Z-2606-2026 | Class II | Medline Convenience kits containing recalled Sw... | May 28, 2026 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.