Automated Impella Controller (AIC); Product Code: 0042-0000-US;

FDA Device Recall #Z-2211-2026 - Class I, April 20, 2026

Recall Summary

Recall Number Z-2211-2026
Classification Class I - Serious risk
Date Initiated April 20, 2026
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Abiomed, Inc.
Location Danvers, MA
Product Type Devices
Quantity 91 units

Product Description

Automated Impella Controller (AIC); Product Code: 0042-0000-US;

Reason for Recall

A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of CA, CT, FL, LA, MA, MI, MO, MS, ND, NY, OH, PA, TX, WY and the countries of Brazil, Canada, Colombia, Hong Kong, Italy, Kuwait, Saudi Arabia, UAE, United Kingdom;

Lot / Code Information

Product Code: 0042-0000-US; UDI-DI: 00813502010022; Serial Numbers: IC2449, IC2451, IC2452, IC2453, IC2459, IC2460, IC1096, IC1142, IC1202, IC1203, IC1224, IC1231, IC1238, IC1284, IC1285, IC1390, IC1545, IC1577, IC1665, IC1708, IC1741, IC1852, IC1853, IC2029, IC2071, IC2154, IC2277, IC2278, IC2314, IC2315, IC2368, IC2369, IC2370, IC2442, IC2443, IC2444, IC2461, IC2471, IC2476, IC2477, IC2478, IC2479, IC2575, IC2989, IC3127, IC3128, IC3129, IC3130, IC3166, IC3167, IC3213, IC3223, IC3306, IC3307, IC3308, IC3309, IC3342, IC3398, IC3475, IC3656, IC3658, IC3663, IC3664, IC3665, IC3666, IC3674, IC3675, IC3719, IC3720, IC3776, IC3812, IC3887, IC3888, IC4045, IC4060, IC1301, IC1544, IC1572, IC2686, IC2687, IC2848, IC3174, IC3269, IC3272, IC3277, IC3278, IC3281, IC3282, IC3284, IC3294, IC3783;

Other Recalls from Abiomed, Inc.

Recall # Classification Product Date
Z-2683-2026 Class I ABIOMED Impella CP Introducer Kit 14 Fr, 13 and... Jun 11, 2026
Z-2689-2026 Class I ABIOMED Impella CP Set, Pump Set Code 0048-0014... Jun 11, 2026
Z-2692-2026 Class I ABIOMED Insertion Kit, Pkgd, JP, Pump Set Code ... Jun 11, 2026
Z-2684-2026 Class I ABIOMED Introducer Kit 23 Fr, 11 cm, Pump Set C... Jun 11, 2026
Z-2682-2026 Class I ABIOMED Impella CP Set, Pump Set Code 0048-0003... Jun 11, 2026

Frequently Asked Questions

Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery, the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.