VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

FDA Device Recall #Z-1942-2026 - Class I, March 25, 2026

Recall Summary

Recall Number Z-1942-2026
Classification Class I - Serious risk
Date Initiated March 25, 2026
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Ventec Life Systems, Inc.
Location Bothell, WA
Product Type Devices
Quantity 42

Product Description

VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

Reason for Recall

Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk

Distribution Pattern

US: OH, GA, CA, MO

Lot / Code Information

UDI-DI (REF/UDI-DI): PRT-01198-000/00855573007914, PRT-01185-000/00855573007877 Package/Device Serial Number (package serial number may include a "P" at the end): 152884, 152885, 152886, 152887, 152888, 152889, 152890, 152891, 152892, 152893, 152895, 152896, 152897, 152898, 152899, 152900, 152901, 152902, 152903, 152904, 152908, 152909, 152910, 152911, 152912, 152913, 152914, 152915, 152916, 152917, 152918, 152919, 152986, 152987, 155236, 156507, 156512, 156515, 156516, 156517, 156524, 156532

Other Recalls from Ventec Life Systems, Inc.

Recall # Classification Product Date
Z-2525-2026 Class II VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pr... May 8, 2026
Z-1196-2025 Class I VOCSN Multi-Function Ventilators: VOCSN+Pro (V... Feb 3, 2025
Z-1070-2024 Class I VOCSN Patient Breathing Package (Pediatric, Act... Jan 9, 2024
Z-1902-2021 Class II Ventec Life Systems VOCSN Multi-Function Ventil... May 10, 2021

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.