Karl Storz NEURO-FIBERSCOPE, Model # 11282BN1, Flexible Pediatric Neuroscope
FDA Device Recall #Z-1530-2020 — Class II — August 27, 2019
Recall Summary
| Recall Number | Z-1530-2020 |
| Classification | Class II — Moderate risk |
| Date Initiated | August 27, 2019 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Karl Storz Endoscopy |
| Location | El Segundo, CA |
| Product Type | Devices |
| Quantity | 73 scopes |
Product Description
Karl Storz NEURO-FIBERSCOPE, Model # 11282BN1, Flexible Pediatric Neuroscope
Reason for Recall
During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK) and tray information, specified for the STERIS System 1E (SS1E) sterilization method within the instruction manual is incorrect. If the user uses SS1E to sterilize the impacted scopes and referred to the manual concerning the QCK information, the sterility of reprocessed scope cannot be guaranteed.
Distribution Pattern
US: AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, MA, MD, MI, MN, MO, NC, NH, NJ, OH, OR, PA, SC, TN, TX, VA, WA, WI, WV, OUS: Canda
Lot / Code Information
All Lots
Other Recalls from Karl Storz Endoscopy
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1062-2025 | Class II | KARL STORZ - ENDOSKOPE, REF: 27001K, Ureterosco... | Dec 19, 2024 |
| Z-1052-2025 | Class II | KARL STORZ - ENDOSKOPE, REF: 27030KA, Cysto-Ure... | Dec 19, 2024 |
| Z-1051-2025 | Class II | KARL STORZ - ENDOSKOPE, REF: 27003L, MICHEL Ure... | Dec 19, 2024 |
| Z-1055-2025 | Class II | KARL STORZ - ENDOSKOPE, REF: 27010L, Uretero-Re... | Dec 19, 2024 |
| Z-1057-2025 | Class II | KARL STORZ - ENDOSKOPE, REF: 27001L, Uretero-Re... | Dec 19, 2024 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.