BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASS...
FDA Recall #Z-0430-2026 — Class I — October 17, 2025
Product Description
BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris Infusion System Inspection Safety Reminder, Part: BD-163124, Revision: 00; Cleaning and Disinfecting Procedure, Part: P00000864; Cleaning and Disinfecting Quick Reference Guide
Reason for Recall
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.
Recalling Firm
CareFusion 303, Inc. — San Diego, CA
Classification
Class I — Reasonable probability of serious adverse health consequences or death.
Product Type
Devices
Product Quantity
2,574,396
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Guam, American Samoa, Northern Mariana Islands; and countries of SG, BH, AE, JP, CA, IT, QA, PH, DE, TW, PK, KR, BW, SA, MY, IL, NZ, AU, IN, GB, CN, NL, FR, ZA, BE, MX, CH, GI, KW, and TR.
Code Information
All Alaris Pump Modules distributed after June 19, 2004. REF/UDI-DI: 8100/10885403810015, 10885403810046, 10885403810039, 10885403222054, 10885403517723, 10885403213212, 10885403211423; 49000437/10885403471018; 49000270/10885403247743
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated