Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;

FDA Recall #Z-2027-2025 — Class II — May 22, 2025

Recall #Z-2027-2025 Date: May 22, 2025 Classification: Class II Status: Ongoing

Product Description

Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;

Reason for Recall

Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.

Recalling Firm

Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

1,625 units (1450 US, 175 OUS)

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Bahamas.

Code Information

Siemens Material Number (SMN): 11099300; UDI-DI: 00630414595771; Lot Number: 0000195412;

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated