Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt...

FDA Recall #Z-1472-2026 — Class II — January 21, 2026

Recall #Z-1472-2026 Date: January 21, 2026 Classification: Class II Status: Ongoing

Product Description

Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure

Reason for Recall

Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.

Recalling Firm

Agiliti Health - Ellis — Ellis, KS

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

4286

Distribution

US Nationwide distribution.

Code Information

UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated