GE Heathcare Giraffe OmniBed Carestation, Model Number 2082844-001-XXX; neo natal incubator

FDA Recall #Z-0048-2025 — Class I — September 3, 2024

Recall #Z-0048-2025 Date: September 3, 2024 Classification: Class I Status: Ongoing

Product Description

GE Heathcare Giraffe OmniBed Carestation, Model Number 2082844-001-XXX; neo natal incubator

Reason for Recall

GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.

Recalling Firm

DATEX--OHMEDA, INC. — Wauwatosa, WI

Classification

Class I — Reasonable probability of serious adverse health consequences or death.

Product Type

Devices

Distribution

Worldwide distribution.

Code Information

All models

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated