POWERSEAL 5 mm, 37 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in gener...

FDA Device Recall #Z-1700-2024 — Class II — March 7, 2024

Recall Summary

Recall Number Z-1700-2024
Classification Class II — Moderate risk
Date Initiated March 7, 2024
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Olympus Corporation of the Americas
Location Center Valley, PA
Product Type Devices
Quantity 850 units: (400 units US);(450 units OUS)

Product Description

POWERSEAL 5 mm, 37 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Model/Catalog Number: PS-0537CJDA

Reason for Recall

Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting in delay in surgery.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of AU, CA, HK, KR.

Lot / Code Information

UDI-DI: 00821925044555 Lot numbers: CA182543, CA182544, CA182546, CA182548, CA191230

Other Recalls from Olympus Corporation of the Americas

Recall # Classification Product Date
Z-1641-2026 Class II Brand Name: Olympus HALO PKS Cutting Forceps, 5... Feb 25, 2026
Z-1640-2026 Class II Brand Name: Olympus PK Cutting Forceps, 5mm 33c... Feb 25, 2026
Z-1642-2026 Class II Brand Name: Olympus PKS Cutting Forceps, 5mm 33... Feb 25, 2026
Z-1639-2026 Class II Brand Name: Olympus Everest Bipolar Cutting For... Feb 25, 2026
Z-1643-2026 Class II Brand Name: Olympus PKS Cutting Forceps w/Cord,... Feb 25, 2026

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.