Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomograp...

FDA Recall #Z-0222-2026 — Class II — September 12, 2025

Recall #Z-0222-2026 Date: September 12, 2025 Classification: Class II Status: Ongoing

Product Description

Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uEXPLORER MD, Rx Only

Reason for Recall

Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

Recalling Firm

UIH Technologies LLC — Houston, TX

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

2 UNITS

Distribution

US: OUS:

Code Information

UDI: (01)06971576832040; Serial Number: F00002, F00015

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated