Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012, Software Ver...
FDA Device Recall #Z-1895-2024 — Class II — May 16, 2024
Recall Summary
| Recall Number | Z-1895-2024 |
| Classification | Class II — Moderate risk |
| Date Initiated | May 16, 2024 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. |
| Location | Best |
| Product Type | Devices |
| Quantity | 1 unit (US) |
Product Description
Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012, Software Version Number 8.1.100.
Reason for Recall
When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
Distribution Pattern
Domestic: OK International: India, Spain, U.A.E.
Lot / Code Information
Product number: 722012, UDI/DI: 00884838059054; Serial Number: 2138.
Other Recalls from PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1346-2026 | Class II | Azurion 7 B12 System Model Numbers:(1) 722067,... | Jan 26, 2026 |
| Z-1350-2026 | Class II | Azurion 5 M12 System Model Numbers: (1)722227,... | Jan 26, 2026 |
| Z-1349-2026 | Class II | Azurion 7 M20 System Model Numbers: (1)722079,... | Jan 26, 2026 |
| Z-1348-2026 | Class II | Azurion 7 M12 System Model Numbers: (1)722078,... | Jan 26, 2026 |
| Z-1345-2026 | Class II | Azurion 3 M15 System Model Numbers: (1) 722064... | Jan 26, 2026 |
Frequently Asked Questions
A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.