Convenience kits; DYNJ901046M GYN ROBOTIC DYNJ909120D ROBOTIC PROSTATE

FDA Recall #Z-0450-2026 — Class I — October 10, 2025

Recall #Z-0450-2026 Date: October 10, 2025 Classification: Class I Status: Ongoing

Product Description

Convenience kits; DYNJ901046M GYN ROBOTIC DYNJ909120D ROBOTIC PROSTATE

Reason for Recall

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

Recalling Firm

Medline Industries, LP — Northfield, IL

Classification

Class I — Reasonable probability of serious adverse health consequences or death.

Product Type

Devices

Product Quantity

96,406 total units

Distribution

International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;

Code Information

GYN ROBOTIC DYNJ901046M UDI-DI 40198459130381 (case), 10198459130380 (ea) Lots 25FBM810 25GBK307; ROBOTIC PROSTATE DYNJ909120D UDI-DI 40198459130541 (case), 10198459130540 (ea) Lots 25GBT821 25IBF809

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated