MINI STICK MAX 5F X 15 CM STIFF .018 NI/TU NON-ECHO 2.75" PG-Used for the percutaneous introduct...

FDA Recall #Z-1347-2024 — Class II — January 30, 2024

Recall #Z-1347-2024 Date: January 30, 2024 Classification: Class II Status: Ongoing

Product Description

MINI STICK MAX 5F X 15 CM STIFF .018 NI/TU NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-776

Reason for Recall

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Recalling Firm

Angiodynamics, Inc. — Queensbury, NY

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

40 units

Distribution

Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.

Code Information

UDI: 15051684023191 UPN: H965457761 Lot Number: 5787840 5803520 5805069

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated