Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro...

FDA Recall #Z-1577-2025 — Class II — March 10, 2025

Recall #Z-1577-2025 Date: March 10, 2025 Classification: Class II Status: Ongoing

Product Description

Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).

Reason for Recall

Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.

Recalling Firm

Maquet Cardiovascular, LLC — Wayne, NJ

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

110 units

Distribution

Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL, GA, IA, ID, IL, IN, KY, MD, MI, MN, MO, NC, NH, NJ, NY, OH, PA, PR, TX, UT, VA, WI, WV and the countries of China, Japan, Netherlands, New Zealand, Singapore, United Kingdom, United States.

Code Information

Model No. VH-3010; UDI-DI: 0060756770826; Serial No. H24080065G, H24080074G, H24080012G, H24080022G, H24080184G, H24080185G, H24080051G, H24080117G, H24080069G, H24080016G, H24080079G, H24080041G, H24080056G, H24080030G, H24080068G, H24080034G, H24080058G, H24080026G, H24080033G, H24080020G, H24080032G, H24080053G, H24080066G, H24080077G, H24080144G, H24080123G, H24080042G, H24080070G, H24080008G, H24080003G, H24080174G, H24080064G, H24080027G, H24080044G, H24080047G, H24080049G, H24080061G, H24080078G, H24080010G, H24080138G, H24080080G, H24080006G, H24080015G, H24080038G, H24080085G, H24080084G, H24080031G, H24080101G, H24080148G, H24080040G, H24080055G, H24080025G, H24080052G, H24080048G, H24080093G, H24080126G, H24080073G, H24080007G, H23060082G, H24080075G, H24080113G, H24080127G, H24080011G, H24080005G, H24080013G, H24080023G, H24080057G, H24080110G, H24080009G, H24080067G, H24080029G, H24080043G, H24080183G, H24080165G, H24080017G, H24080001G, H24080037G, H24080060G, H24080072G, H24080076G, H24080045G, H24080024G, H24080028G, H24080114G, H24080136G, H24080062G, H24080018G, H24080054G, H24080063G, H24080112G, H24080118G, H24080071G, H24080182G, H24080002G, H24080004G, H24080014G, H24080019G, H24080021G, H24080035G, H24080036G, H24080039G, H24080046G, H24080050G, H24080059G, H24080081G, H24080082G, H24080083G, H24080125G, H24080186G, H24080189G.

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated