Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 530 ...

FDA Recall #Z-0217-2026 — Class II — September 12, 2025

Recall #Z-0217-2026 Date: September 12, 2025 Classification: Class II Status: Ongoing

Product Description

Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 530 MD, Rx Only

Reason for Recall

Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

Recalling Firm

UIH Technologies LLC — Houston, TX

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

8 units

Distribution

US: OUS:

Code Information

Serial Numbers: ; UDI: (01)06971576831012/ Serial Numbers: 300162 300163 300174 300176 304002 304003 304004 304006

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated