MINI STICK MAX 4F X 10 CM STIFF .018 NI/TU NON-ECHO 2.75" PG- Used for the percutaneous introduc...

FDA Recall #Z-1342-2024 — Class II — January 30, 2024

Recall #Z-1342-2024 Date: January 30, 2024 Classification: Class II Status: Ongoing

Product Description

MINI STICK MAX 4F X 10 CM STIFF .018 NI/TU NON-ECHO 2.75" PG- Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-762

Reason for Recall

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Recalling Firm

Angiodynamics, Inc. — Queensbury, NY

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

600 units

Distribution

Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.

Code Information

UDI: 15051684023054 UPN: H965457621 Lot Number: Lot: 5788541 5789982 5791594 5794608 5797468 5798845

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated