Technopath Multichem S Plus Level 2 - Product Usage: Multichem S Plus (Assayed) is intended for u...

FDA Device Recall #Z-1363-2019 — Class III — April 9, 2019

Recall Summary

Recall Number Z-1363-2019
Classification Class III — Low risk
Date Initiated April 9, 2019
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Abbott Laboratories
Location Abbott Park, IL
Product Type Devices
Quantity 1 unit

Product Description

Technopath Multichem S Plus Level 2 - Product Usage: Multichem S Plus (Assayed) is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.

Reason for Recall

Devices were delivered without the required dry ice.

Distribution Pattern

Distribution to Latvia.

Lot / Code Information

LN 5P78-11; Lot 18005182

Other Recalls from Abbott Laboratories

Recall # Classification Product Date
Z-0961-2026 Class II Alinity ci-series System Control Module (SCM), ... Nov 13, 2025
Z-0489-2026 Class II Brand Name: Alinity hq Analyzer Product Name: ... Sep 18, 2025
Z-1652-2025 Class II Brand Name: CELL-DYN Ruby Product Name: CELL-D... Mar 18, 2025
Z-1655-2025 Class II Brand Name: CELL-DYN Ruby Product Name: CELL-D... Mar 18, 2025
Z-1653-2025 Class II Brand Name: CELL-DYN Ruby Product Name: CELL-D... Mar 18, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.