GREER DIA - KIT ANTIGEN M. FAENI, 2.0mL 20 MG/ML, Analyte Specific Reagent, Item: K62, HYPERSENSI...

FDA Device Recall #Z-0348-2023 — Class II — October 25, 2022

Recall Summary

Recall Number Z-0348-2023
Classification Class II — Moderate risk
Date Initiated October 25, 2022
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Greer Laboratories, Inc.
Location Lenoir, NC
Product Type Devices
Quantity 102 units

Product Description

GREER DIA - KIT ANTIGEN M. FAENI, 2.0mL 20 MG/ML, Analyte Specific Reagent, Item: K62, HYPERSENSITIVITY PNEUMONITIS test system, gel diffusion kit for demonstrating precipitating antibodies.

Reason for Recall

The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative result interpretation. The issue was identified by the firm after failing to meet lot requirements for potency. The lot release error was discovered on October 08, 2021, after a failed potency test was determined during 6-month stability evaluation of retention material for the lots. The subsequent manufacturing investigation, which included a review of the lot release potency documentation, found that the initial lot release potency test for the lot was misinterpreted.

Distribution Pattern

US Distribution to states of: CA, CO, IA, KY, MA, MD, MO, NC, PA, UT, VA, and International to countries of: Canada, Denmark

Lot / Code Information

Lot Number 389592, exp 6/12/2023

Other Recalls from Greer Laboratories, Inc.

Recall # Classification Product Date
Z-0347-2023 Class II GREER M. FAENI, 0.25 mL 20 MG/ML, Item: K6, Ana... Oct 25, 2022

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.