Leica Surgical Operating Microscopes, PROvido Part Numbers: a. 10448976, b. 10448950

FDA Device Recall #Z-1851-2022 — Class II — August 26, 2022

Recall Summary

Recall Number Z-1851-2022
Classification Class II — Moderate risk
Date Initiated August 26, 2022
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Leica Microsystems, Inc.
Location deerfield, IL
Product Type Devices
Quantity 32 devices

Product Description

Leica Surgical Operating Microscopes, PROvido Part Numbers: a. 10448976, b. 10448950

Reason for Recall

During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.

Distribution Pattern

US Nationwide distribution.

Lot / Code Information

a. 10448976, Serial Numbers: 120322001, 300122001, 170921001, 301221001, 40122001, 140522001, 151221002, 301021002, 81221002, 100222002, 40222002, 220821001, 300122002, 110322001, 50322001, 180921001, 260821001, 90622001, 120422001, 100522001, 90522001, 210821001, 250122001, 301021001, 100222001; b. 10448950, Serial Numbers: 160222001, 90721001, 50422001, 50422002, 91221001, 160921001, 61221001

Other Recalls from Leica Microsystems, Inc.

Recall # Classification Product Date
Z-0477-2023 Class II Leica MICROSYSTEMS M530 OHX, Part Number 104487... Nov 21, 2022
Z-0476-2023 Class II Leica MICROSYSTEMS Provido, Part Numbers 104489... Nov 21, 2022
Z-0475-2023 Class II Leica MICROSYSTEMS Proveo 8, Part Number 104490... Nov 21, 2022
Z-1849-2022 Class II Leica Surgical Operating Microscopes, Arveo, Pa... Aug 26, 2022
Z-1848-2022 Class II Leica Surgical Operating Microscopes, ARveo 8, ... Aug 26, 2022

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.