BD Trucount Tubes (Cat. No. 663028), used for determining absolute counts of leukocytes in blood....
FDA Device Recall #Z-0172-2023 — Class II — October 3, 2022
Recall Summary
| Recall Number | Z-0172-2023 |
| Classification | Class II — Moderate risk |
| Date Initiated | October 3, 2022 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Becton, Dickinson and Company, BD Biosciences |
| Location | San Jose, CA |
| Product Type | Devices |
| Quantity | 11,709 tubes (corrected 11/04/2022) |
Product Description
BD Trucount Tubes (Cat. No. 663028), used for determining absolute counts of leukocytes in blood. BD Trucount Tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount Tubes can be used with the BD FACS Loader.
Reason for Recall
Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.
Distribution Pattern
U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.: Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Korea, Mexico, Malaysia, New Zealand, Panama, Peru, Philippines, Paraguay, Singapore, Thailand, Taiwan, and Uruguay
Lot / Code Information
Corrected as of 11/04/2022: Catalog Number: 663028 UDI-DI Code: 00382906630285 Lot Numbers: 2031238 2063994 2095615 2145791 2221697 2237045
Other Recalls from Becton, Dickinson and Company, BD Bio...
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2351-2024 | Class II | BD Multitest 6-Color TBNK CE-IVD, REF: 644611; ... | May 8, 2024 |
| Z-0641-2024 | Class II | CD11b APC: ASR, REF: 340936, and CE, REF: 333143 | Oct 25, 2023 |
| Z-0173-2023 | Class II | BD Multitest 6-Color TBNK with optional BD Truc... | Oct 3, 2022 |
| Z-0171-2023 | Class II | BD Trucount Tubes (Cat. No. 340334), used for d... | Oct 3, 2022 |
| Z-1215-2022 | Class II | BD Anti-Kappa PE (Cat. No. 346587) is an analyt... | Mar 25, 2022 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.