Hamilton-MR1, REF: 161010, Intensive Care Ventilator

FDA Device Recall #Z-2528-2023 — Class I — July 26, 2023

Recall Summary

Recall Number Z-2528-2023
Classification Class I — Serious risk
Date Initiated July 26, 2023
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Hamilton Medical, Inc.
Location Reno, NV
Product Type Devices
Quantity 12 units

Product Description

Hamilton-MR1, REF: 161010, Intensive Care Ventilator

Reason for Recall

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Distribution Pattern

Distribution US nationwide, including Puerto Rico. OUS: Not provided

Lot / Code Information

UDI: 07630002800761 & 07630002813556 / Affected Serial Numbers: 3422 3423 3424 3425 3426 3427 3428 3429 3469 3503 3510 3800

Other Recalls from Hamilton Medical, Inc.

Recall # Classification Product Date
Z-2530-2023 Class I Hamilton-T1, REF: MSP161339/02, ESM Shielding S... Jul 26, 2023
Z-2525-2023 Class I Hamilton-C1, REF: 161001, Portable Intensive Ca... Jul 26, 2023
Z-2529-2023 Class I Hamilton C1/T1/MR1, REF: MSP161502/12, Control ... Jul 26, 2023
Z-2526-2023 Class I Hamilton-T1, REF: 161006, Intensive Care and Tr... Jul 26, 2023
Z-2527-2023 Class I Hamilton-T1, REF: 161009, Intensive Care and Tr... Jul 26, 2023

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.