LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTE...

FDA Device Recall #Z-1840-2019 — Class II — April 23, 2019

Recall Summary

Recall Number Z-1840-2019
Classification Class II — Moderate risk
Date Initiated April 23, 2019
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm CooperSurgical, Inc.
Location Trumbull, CT
Product Type Devices
Quantity 19

Product Description

LEEP Precision Integrated System 120V, Model LP-10-120 Product Usage: The LEEP PRECISION INTEGRATED SYSTEM l 20V combines three components - LEEP PRECISION Cart, LEEP PRECISION Generator, and the LEEP PRECISION Smoke Evacuator for loop electrosurgical excision procedures. The loop electrosurgical excision procedure (LEEP) utilizes a very thin wire in the shape of a loop heated by electrical current generated by a power supply to efficiently remove abnormal cells. The LEEP PRECISION INTEGRATED SYSTEM has been designed to provide an efficient workstation that can efficiently meet all the unique LEEP requirements of the clinic and practice-based facilities .

Reason for Recall

2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses.

Distribution Pattern

US Nationwide Distribution in the states of CA, CT, IN, MA, MI, MN, NC, NV, NY, OH, TX, WA and WI.

Lot / Code Information

2018110001,2018110002, 2018110003, 2018110004, 2018110005,2018110006, 2018110007, 2018110008,2018110011, 2018110012,2018110013, 2018110014, 2018110015,2018110016, 2018110017, 2018110019, 2018110020,2018110021,& 2018110022

Other Recalls from CooperSurgical, Inc.

Recall # Classification Product Date
Z-2136-2025 Class II Endosee System Convenience Kit. Model Number: E... Jun 11, 2025
Z-2135-2025 Class II HSG Procedure Tray. Model Number: 6050T. Th... Jun 11, 2025
Z-2137-2025 Class II Endosee System Convenience Kit with IV Tube. Mo... Jun 11, 2025
Z-0687-2025 Class II Brand Name: K-Systems Product Name: G73 Dry Ba... Nov 15, 2024
Z-0688-2025 Class II Brand Name: K-Systems Product Name: G85 Mini I... Nov 15, 2024

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.