REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only w...

FDA Device Recall #Z-0612-2020 — Class II — August 16, 2019

Recall Summary

Recall Number Z-0612-2020
Classification Class II — Moderate risk
Date Initiated August 16, 2019
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Enztec Limited
Location Christchurch
Product Type Devices
Quantity 312 Femoral Impactors

Product Description

REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE

Reason for Recall

The femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument. This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response that may require revision surgery.

Distribution Pattern

CA

Lot / Code Information

Lot number Distribution Date: 161471 08-Jun-16, 161472 08-Jun-16, 161473 08-Jun-16, 161474 08-Jun-16, 161475 08-Jun-16, 161476 08-Jun-16, 161477 08-Jun-16, 161756 29-Jun-16, 161757 29-Jun-16, 161758 29-Jun-16, 161759 29-Jun-16, 161760 29-Jun-16, 161761 29-Jun-16, 161762 29-Jun-16, 161822 22-Jul-16, 161823 22-Jul-16, 161824 22-Jul-16, 161825 22-Jul-16, 161826 22-Jul-16, 161827 22-Jul-16, 161828 22-Jul-16, 170760 24-Mar-17, 170761 24-Mar-17, 170762 24-Mar-17, 170763 24-Mar-17, 170764 24-Mar-17, 170765 24-Mar-17, 170766 24-Mar-17, 170767 24-Mar-17, 170768 24-Mar-17, 171989 23-Jun-17 ,

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.