ADVIA Chemistry Enzymatic Creatinine (ECRE_2) reagent, SMN 10335869

FDA Recall #Z-1230-2020 — Class II — December 18, 2019

Recall #Z-1230-2020 Date: December 18, 2019 Classification: Class II Status: Terminated

Product Description

ADVIA Chemistry Enzymatic Creatinine (ECRE_2) reagent, SMN 10335869

Reason for Recall

There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.

Recalling Firm

Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

1,015 units

Distribution

US Nationwide

Code Information

UDI Number: 00630414526690 Lots: All lots

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated