ABL80-FLEX CO-OX 393-841 Box Label, REF 393-841, IVD, CE, UDI: 05700693938417 - Product Usage: I...
FDA Device Recall #Z-2468-2020 — Class II — May 11, 2020
Recall Summary
| Recall Number | Z-2468-2020 |
| Classification | Class II — Moderate risk |
| Date Initiated | May 11, 2020 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Radiometer Medical ApS |
| Location | Bronshoj, N/A |
| Product Type | Devices |
| Quantity | 1981 |
Product Description
ABL80-FLEX CO-OX 393-841 Box Label, REF 393-841, IVD, CE, UDI: 05700693938417 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.
Reason for Recall
Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accession number. Misinterpretation of a barcode, including e.g. patient ID or accession number, may result in patient mix up and/or delayed medical treatment.
Distribution Pattern
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AK, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CA.
Lot / Code Information
All lots
Other Recalls from Radiometer Medical ApS
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1659-2025 | Class II | Radiometer ABL90 Series - Radiometer ABL90 FLEX... | Mar 21, 2025 |
| Z-2426-2024 | Class II | ABL800 FLEX Model Numbers 393-800 and 393-801 ... | May 31, 2024 |
| Z-1394-2023 | Class II | AQURE REF 933-599 Software Versions 2.5.2 2... | Mar 8, 2023 |
| Z-1321-2022 | Class II | ABL800 Flex Analyzer Model Numbers 393-800 and ... | May 18, 2022 |
| Z-1115-2022 | Class II | Radiometer, REF:956-552, PIC050- Arterial Blood... | Apr 8, 2022 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.