Kangaroo Joey Safety Screw Spike with Flush Bag, Anti-free Flow,1000 mL, Product Code 765100
FDA Recall #Z-2100-2020 — Class II — December 26, 2019
Product Description
Kangaroo Joey Safety Screw Spike with Flush Bag, Anti-free Flow,1000 mL, Product Code 765100
Reason for Recall
The feeding spike sets may leak at the interface of the tube and spike connector.
Recalling Firm
Cardinal Health Inc. — Dublin, OH
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
114,731, 290 total units
Distribution
Worldwide distribution.
Code Information
All
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated