ARTIS Pheno-Interventional Fluoroscopic X-Ray System Model Number: 10849000 - Product Usage:AXIO...
FDA Device Recall #Z-1340-2020 — Class II — January 24, 2020
Recall Summary
| Recall Number | Z-1340-2020 |
| Classification | Class II — Moderate risk |
| Date Initiated | January 24, 2020 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Siemens Medical Solutions USA, Inc |
| Location | Malvern, PA |
| Product Type | Devices |
| Quantity | 96 US |
Product Description
ARTIS Pheno-Interventional Fluoroscopic X-Ray System Model Number: 10849000 - Product Usage:AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Axiom Artis can also support the acquisition of position triggered imaging for spatial data synthesis.
Reason for Recall
Potential exists for one or more axes of the robotic stand to become blocked by their safety brakes; In the event the axis is blocked by its safety brake, the robotic stand can no longer perform the intended movements
Distribution Pattern
US Nationwide distribution.
Lot / Code Information
All serial numbers
Other Recalls from Siemens Medical Solutions USA, Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1175-2026 | Class II | Interventional Fluoroscopic X-Ray System: LUMIN... | Dec 29, 2025 |
| Z-1174-2026 | Class II | Interventional Fluoroscopic X-Ray System: LUMIN... | Dec 29, 2025 |
| Z-1241-2026 | Class II | NAEOTOM Alpha.Prime Software applications: s... | Dec 19, 2025 |
| Z-1240-2026 | Class II | NAEOTOM Alpha Software applications: syngo.C... | Dec 19, 2025 |
| Z-1239-2026 | Class II | SOMATOM X.ceed Software applications: syngo.... | Dec 19, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.