SIGNA Artist, Nuclear Magnetic Resonance Imaging System
FDA Device Recall #Z-2061-2021 — Class II — June 4, 2021
Recall Summary
| Recall Number | Z-2061-2021 |
| Classification | Class II — Moderate risk |
| Date Initiated | June 4, 2021 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | GE Healthcare, LLC |
| Location | Waukesha, WI |
| Product Type | Devices |
| Quantity | 562 devices |
Product Description
SIGNA Artist, Nuclear Magnetic Resonance Imaging System
Reason for Recall
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.
Distribution Pattern
Worldwide distributions.
Other Recalls from GE Healthcare, LLC
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1566-2022 | Class II | (1) GE Centricity Universal Viewer 6.0, Model 2... | Jul 15, 2022 |
| Z-1430-2022 | Class II | SIGNA Premier magnetic resonance scanner, model... | Jun 24, 2022 |
| Z-1466-2022 | Class II | Centricity PACS Software Version 7.0 SP0.0.4.7 | Jun 14, 2022 |
| Z-1333-2022 | Class II | GE Centricity Universal Viewer Zero Footprint. ... | Jun 8, 2022 |
| Z-1511-2022 | Class II | Service manuals for the CT, PET, or NM table as... | Jun 2, 2022 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.