Computed Tomography X-ray System, Model: uCT 550, GTIN: 06971576831036 Intended use The uCT C...
FDA Device Recall #Z-2641-2020 — Class II — May 23, 2020
Recall Summary
| Recall Number | Z-2641-2020 |
| Classification | Class II — Moderate risk |
| Date Initiated | May 23, 2020 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Shanghai United Imaging Healthcare Co., Ltd. |
| Location | Shanghai, N/A |
| Product Type | Devices |
| Quantity | 3 |
Product Description
Computed Tomography X-ray System, Model: uCT 550, GTIN: 06971576831036 Intended use The uCT Computed Tomography X-ray System uCT 530/uCT 550 is a computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck and vascular).
Reason for Recall
Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.
Distribution Pattern
US Distribution to state of: TX
Lot / Code Information
Serial Numbers: 353020, 353021, and 353023 with software version R001.11.0.729741-Re-1100
Other Recalls from Shanghai United Imaging Healthcare Co...
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1353-2023 | Class II | Shanghai United Imaging Healthcare Co., Ltd. Po... | Mar 1, 2023 |
| Z-0087-2023 | Class II | Positron Emission Tomography and Computed Tomog... | May 30, 2022 |
| Z-2594-2021 | Class II | uEXPLORER PET/CT System - Product Usage: intend... | Aug 2, 2021 |
| Z-2593-2021 | Class II | uMI 550 System - Product Usage: intended to be ... | Aug 2, 2021 |
| Z-1160-2021 | Class II | Digitial Medica X-ray Imagining System, Model: ... | Jan 25, 2021 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.