Pencan Spinal Needle Tray. Product Catalog Number: 333865. Local anesthesia of the skin prior...
FDA Device Recall #Z-0569-2018 — Class II — June 21, 2017
Recall Summary
| Recall Number | Z-0569-2018 |
| Classification | Class II — Moderate risk |
| Date Initiated | June 21, 2017 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | B. Braun Medical, Inc. |
| Location | Allentown, PA |
| Product Type | Devices |
| Quantity | 11,920 units |
Product Description
Pencan Spinal Needle Tray. Product Catalog Number: 333865. Local anesthesia of the skin prior to insertion of the regional anesthesia needle.
Reason for Recall
B. Braun Medical Inc. (BBMI) initiated this recall of various BBMI convenience kits containing the 22GA 1-1/2in SafetyGlide Needle due to the presence of loose polypropylene foreign matter was above release specification. This foreign matter has been identified as a product hub material which has been tested for biocompatibility per ISO 10993 during the product development process.
Distribution Pattern
US Nationwide Distribution.
Lot / Code Information
0061483775. 0061486469. 0061519983. 0061525310. 0061529050. 0061531352. 0061535252. 0061540004.
Other Recalls from B. Braun Medical, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1021-2025 | Class II | Design Options COMBINED SPINAL/EPIDURAL TRAY- ... | Dec 4, 2024 |
| Z-3144-2024 | Class II | AET36 ANESTH EXT SET - Anesthesia Extension Set... | Aug 8, 2024 |
| Z-2387-2024 | Class II | Infusomat SPACE PUMP IV SET, 2 CARESITES, ASV-U... | Jun 17, 2024 |
| Z-2380-2024 | Class II | Infusomat 60DROP METRISET PUMP SET, 3 SAFELINE... | Jun 17, 2024 |
| Z-2391-2024 | Class II | Infusomat UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE... | Jun 17, 2024 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.