AXIOM Artis dBC, Model Number 7728392
FDA Device Recall #Z-2027-2019 — Class II — June 12, 2019
Recall Summary
| Recall Number | Z-2027-2019 |
| Classification | Class II — Moderate risk |
| Date Initiated | June 12, 2019 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Siemens Medical Solutions USA, Inc |
| Location | Malvern, PA |
| Product Type | Devices |
| Quantity | 53 |
Product Description
AXIOM Artis dBC, Model Number 7728392
Reason for Recall
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
Distribution Pattern
US Nationwide Distribution
Lot / Code Information
Serial Numbers: 44125 44151 44101 44246 44185 44262 44263 44199 44105 44084 44080 44072 44243 44198 44171 44275 44215 44255 44081 44242 44265 44189 44289 44145 44060 44144 44184 44286 44223 44068 44193 44244 44201 44251 44229 44304 44152 44136 44254 44277 44279 44268 44206 44278 44259 44260 44168 44267 44110 44218 44165 44271 44135
Other Recalls from Siemens Medical Solutions USA, Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1175-2026 | Class II | Interventional Fluoroscopic X-Ray System: LUMIN... | Dec 29, 2025 |
| Z-1174-2026 | Class II | Interventional Fluoroscopic X-Ray System: LUMIN... | Dec 29, 2025 |
| Z-1241-2026 | Class II | NAEOTOM Alpha.Prime Software applications: s... | Dec 19, 2025 |
| Z-1240-2026 | Class II | NAEOTOM Alpha Software applications: syngo.C... | Dec 19, 2025 |
| Z-1239-2026 | Class II | SOMATOM X.ceed Software applications: syngo.... | Dec 19, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.