Stryker ACM (Advanced Cement Mixing): 180-Gram Cement Cartridge with Breakaway Femoral Nozzle-All...
FDA Recall #Z-2597-2023 — Class II — August 7, 2023
Product Description
Stryker ACM (Advanced Cement Mixing): 180-Gram Cement Cartridge with Breakaway Femoral Nozzle-Allows the user to mix bone cement with a constant high vacuum through the process Catalog Number: 0206-530-000
Reason for Recall
Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site
Recalling Firm
Stryker Corporation — Portage, MI
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
224 units US; 718 units OUS
Distribution
Nationwide Foreign: Austria Belgium Canada France Germany Hong Kong Italy Malaysia Poland Singapore South Korea Spain Taiwan Thailand UK
Code Information
GTIN: 04546540857880 Lot Numbers: 22332012 22336012 22333012 22355012 23052012
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated