The PREVI¿ Color Gram dyes are used with the PREVI¿ Color Gram instrument to stain bacterial and ...
FDA Device Recall #Z-2170-2018 — Class II — March 23, 2018
Recall Summary
| Recall Number | Z-2170-2018 |
| Classification | Class II — Moderate risk |
| Date Initiated | March 23, 2018 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | BioMerieux SA |
| Location | Marcy L'Etoile, N/A |
| Product Type | Devices |
| Quantity | FUCHSIN-A (ref: 29522) 15 eaches; IODINE-B (ref: 29523) 10 eaches ; CRYSTAL VIOLET-C (ref: 29524) 10 eaches |
Product Description
The PREVI¿ Color Gram dyes are used with the PREVI¿ Color Gram instrument to stain bacterial and fungal microorganisms for direct examination, according to Gram s method. The PREVI¿ Color Gram instrument sprays the PREVI¿ Color Gram reagents onto microscope slides which have been smeared with specimens to be examined. The staining procedure is rapid and precise. Each reagent has a specific position (A, B, C, D, E) on the PREVI¿ Color Gram platform that is also indicated on the reagent label. An automated slide stainer is a device used to stain histology, cytology, and hematology slides for diagnosis.
Reason for Recall
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
Distribution Pattern
International distribution.
Lot / Code Information
Reagents for use with the PREVI¿ Color Gram: IODINE-B (ref: 29523) Lot number: G59407 Exp. 15-Aug-18 CRYSTAL VIOLET-C (ref: 29524) Lot number: G52306 Exp.19-Jul-18 FUCHSIN-A (ref: 29522): Lot Number Lot number: F89544 Exp. 13-Apr-18
Other Recalls from BioMerieux SA
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0084-2022 | Class II | MYLA Versions: 4.7, 4.7.1, 4.8, 4.8.1 and 4.8.2... | Sep 2, 2021 |
| Z-2225-2021 | Class II | MYLA software. Used to manage microbiology tes... | Jun 22, 2021 |
| Z-0218-2019 | Class II | NucliSENS¿ Magnetic Extraction Reagents | Sep 17, 2018 |
| Z-2331-2018 | Class II | VIDAS FSH, Ref 30407-01, IVD, Rx. The firm nam... | Apr 9, 2018 |
| Z-2169-2018 | Class II | VIDAS TOXO IgG Avidity is an automated qualitat... | Mar 23, 2018 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.