INFX-8000V Fluoroscopic X-Ray Systems; INFINIX CF-I/BP PLUS CARDIAC SYSTEM

FDA Recall #Z-0844-2018 — Class II — December 29, 2017

Recall #Z-0844-2018 Date: December 29, 2017 Classification: Class II Status: Terminated

Product Description

INFX-8000V Fluoroscopic X-Ray Systems; INFINIX CF-I/BP PLUS CARDIAC SYSTEM

Reason for Recall

During an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error.

Recalling Firm

Toshiba American Medical Systems Inc — Tustin, CA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

29 units

Distribution

United States

Code Information

Serial Numbers: UFD13X2015, UFD13X2016, UUC13Z2011, UEC1422013, UFD1442017, UFD1452018, AAA1642005, UUC1452012, UTC14Y2008, ACA1472001, ACA1492002, AAA1492001, AAA14X2002, A8A1532001, A9A1542001, AAA1542003, A9A1582002, ABB1627001, ACB1622003, A9A1662003, AAA1692006, A9A16X2004, A8B16Y2002, A7B16Y2001, ACB16Y2004, AAA1632004, AAA1712007, W4B1622031, W4B14Y2023

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated