Spacelabs Healthcare Arkon Anesthesia Delivery System is intended for use in the hospital environ...

FDA Device Recall #Z-2093-2017 — Class II — April 14, 2017

Recall Summary

Recall Number Z-2093-2017
Classification Class II — Moderate risk
Date Initiated April 14, 2017
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Del Mar Reynolds Medical, Ltd.
Location Hertford
Product Type Devices
Quantity 62 total (US= 19 devices and 32 kits; International = 11 devices)

Product Description

Spacelabs Healthcare Arkon Anesthesia Delivery System is intended for use in the hospital environment and operating room. Absorber Adapter Assembly Service Kit, part number: 050-9045-00. Absorber Adapter Assembly, part number: 650-1648-01. Arkon Absorber Canister Carrier, part number: 380-0916-01.

Reason for Recall

Reports of the Canister Carrier breaking during absorbent exchange or during anesthesia machine transit, causing a leak in the breathing circuit, which could cause hypoventilation in the patient.

Distribution Pattern

US distribution in FL, LA, MS, NC, NC, SC and WY and Internationally to Panama and United Kingdom.

Lot / Code Information

*********the serial numbers and part numbers of units distributed in the US ********* Absorber Adapter Assembly Service Kit part number: 050-9045-00. Devices serial numbers: ARKN-000395, ARKN-000396, ARKN-000397, ARKN-000398, ARKN-000399, ARKN-000400, ARKN-000401, ARKN-000402, ARKN-000403, ARKN-000404, ARKN-000405, ARKN-000406, ARKN-000407, ARKN-000408, ARKN-000409, ARKN-000410, ARKN-000411, ARKN-000412, and ARKN-000413. *********the serial numbers of units distributed international ************************** ARKN-000375, ARKN-000414, ARKN-000415, ARKN-000416, ARKN-000417, ARKN-000418, ARKN-000419, ARKN-000420, ARKN-000421, ARKN-000422, and ARKN-000423.

Other Recalls from Del Mar Reynolds Medical, Ltd.

Recall # Classification Product Date
Z-1460-2017 Class II Arkon Anesthesia Delivery System, Model 99999. ... Feb 8, 2017
Z-1458-2015 Class II Spacelabs Pediatric Flow Sensor Kit, PN: 376-05... Mar 5, 2015
Z-1442-2014 Class I SpaceLabs Arkon Anesthesia Delivery System with... Mar 10, 2014
Z-0352-2014 Class I CAS I/II Absorbers that are parts of the Blease... Oct 17, 2013
Z-1771-2013 Class II Spacelabs ARKON Anesthesia Workstation. Spa... Jun 7, 2013

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.