CONCEPT HEATWAVE Electrode 130 mm x 2.3 mm diameter, REF/Catalog Number CS-023, STERILE, Rx Only ...
FDA Device Recall #Z-0085-2017 — Class II — August 29, 2016
Recall Summary
| Recall Number | Z-0085-2017 |
| Classification | Class II — Moderate risk |
| Date Initiated | August 29, 2016 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | ConMed Corporation |
| Location | Utica, NY |
| Product Type | Devices |
| Quantity | Domestic: 1740; Foreign: 4770 |
Product Description
CONCEPT HEATWAVE Electrode 130 mm x 2.3 mm diameter, REF/Catalog Number CS-023, STERILE, Rx Only Product Usage: The HeatWave Electrode is intended to be used for electro-coagulation of soft tissue in shoulder, ankle, wrist, elbow and knee arthroscopic procedures.
Reason for Recall
CONMED received complaints of damage to the CS-023 electrodes shaft insulation which could result in burn marks on the shaft. If the shaft insulation is damaged in any way, it could increase the risk of patient burn.
Distribution Pattern
Worldwide Distribution - US Nationwide in the states of: CA, HI, IL, IN, KS, LA, MA, NC, NM, NY, OH, OK, PA, TN, TX, UT, WA and WI; plus Foreign distribution to Australia, Canada, France, Germany, Italy, Spain, Czech Republic, Lithuania, Netherlands, Poland, Korea,Morocco, Malaysia, Mexico, Taiwan, and South Africa.
Lot / Code Information
All lots codes manufactured from 09/01/2011 through 11/17/2015: 267221, 301001, 309889, 318652, 319649, 325019, 354759, 334627, 336041, 341704, 343927, 349952, 384294, 354768, 365239, 369773, 374480, 378521, 406137, 385695, 388538, 392382, 394178, 396542, 436808, 413702, 456341, 423877, 427704, 433503, 478513, 451923, 488101, 464062, 468012, 471297, 511327, 482366, 522335, 494840, 499981, 505321, 544240, 515060, 551152, 527734, 531802, 536700, 581072, 550988, 590037, 557461, 570449, 574167, 611804, 588892, 621304, 597982, 598682, 605515, 643396, 619686, 656909, 629202, 634911, 643395, 676005, 654991, 692617, 667118, 671471, 674410, 683144, 695062, 695063, 297982.
Other Recalls from ConMed Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0474-2023 | Class II | Foot Controlled Electrosurgical Suction Coagula... | Nov 18, 2022 |
| Z-1498-2021 | Class II | INFINITY_TRAY Custom Procedural Kit - Canada. ... | Mar 26, 2021 |
| Z-1499-2021 | Class II | KIT_INFINITY_BASE Custom Procedural Kit - Italy... | Mar 26, 2021 |
| Z-0650-2021 | Class II | Infinity ACL Tibial Elbow Guide-indicated for u... | Nov 12, 2020 |
| Z-0651-2021 | Class II | Infinity ACL Tibial Tip Guide- indicated for u... | Nov 12, 2020 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.