Rebstock Holding rod for fixation -For holding and fixation of flexible Leyla Arms and spatula A...

FDA Recall #Z-1184-2024 — Class II — December 21, 2023

Recall #Z-1184-2024 Date: December 21, 2023 Classification: Class II Status: Ongoing

Product Description

Rebstock Holding rod for fixation -For holding and fixation of flexible Leyla Arms and spatula Art. No. 06-06-285 and 06-06-285MOD. surgical instrument.

Reason for Recall

Holding arm could move (twist) uncontrollably in the front area. This could lead to a serious deterioration of patient's safety if the movement is transferred to the spatula, which is attached to the holding arm.

Recalling Firm

Medagent Inc — Portsmouth, NH

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

84 units

Distribution

TX Foreign: Algeria China Dominican Republic Germany India Japan Philippines UK

Code Information

Batches: 6221672, 6222079, 6230282, 6230985, 6230986, 6231808, 6222559, 6231989, 6221673

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated