Dimension Vista¿ Mass CKMB Isoenzyme Calibrator (MMB CAL) an in vitro diagnostic product for the ...

FDA Device Recall #Z-0024-2015 — Class II — August 28, 2014

Recall Summary

Recall Number Z-0024-2015
Classification Class II — Moderate risk
Date Initiated August 28, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Siemens Healthcare Diagnostics, Inc.
Location Newark, DE
Product Type Devices
Quantity 6562

Product Description

Dimension Vista¿ Mass CKMB Isoenzyme Calibrator (MMB CAL) an in vitro diagnostic product for the calibration of Creatine Kinase MB Isoenzyme (MMB) method on the Dimension Vista¿ System.

Reason for Recall

Siemens Healthcare Diagnostics has determined that Dimension Vista® Mass CKMB Isoenzyme Calibrator (MMB CAL) shows drift during routine stability monitoring that exceeds our acceptance criteria for drift over the shelf life of this product. Siemens has observed drift from 3% to 12% at MMB concentrations across the assay range. Depending on quality control limits, this drift may not have been d

Distribution Pattern

Worldwided distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Canada, Denmark, Germany, Finland, France, Italy, Japan, Netherlands, New Zealand, Portugal, Saudi Arabia, Slovenia, South Korea, Spain, and Switzerland.

Lot / Code Information

lot 3JD045, exp. 2014-10-01; lot 4BD064, exp. 2015-02-01; lot 4CD012, exp. 2015-03-01; and lot 4DD065, exp. 2015-05-01.

Other Recalls from Siemens Healthcare Diagnostics, Inc.

Recall # Classification Product Date
Z-1484-2026 Class II Atellica CH Urine Albumin (UAlb). Material Numb... Jan 20, 2026
Z-0491-2026 Class II IMMULITE 2000 Oak Mix Specific Allergen; Sieme... Oct 6, 2025
Z-0492-2026 Class II 3gAllergy Specific IgE Universal Kit; Kit Comp... Oct 6, 2025
Z-0493-2026 Class II Food Panel 3 Specific Allergen; Siemens Materi... Oct 6, 2025
Z-0366-2026 Class II Assay: IMMULITE 2000 Intact PTH; Test Code: iP... Sep 22, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.