PKG, AXIAL HANDLE, W/RATCHET, INSULATED SHAFT, P/N 0250080235. Laparoscopic Manual Instruments ...

FDA Device Recall #Z-0625-2015 — Class II — November 17, 2014

Recall Summary

Recall Number Z-0625-2015
Classification Class II — Moderate risk
Date Initiated November 17, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Stryker Endoscopy
Location San Jose, CA
Product Type Devices
Quantity 180,573 units total

Product Description

PKG, AXIAL HANDLE, W/RATCHET, INSULATED SHAFT, P/N 0250080235. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason for Recall

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Distribution Pattern

Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Lot / Code Information

41811 140622 140726 140727 140729 342723 345252 544042 544043 644785 645845 645850 645852 645854 648085 06CON06065 744231 744232 745522 746745 840675 842159 842161 842197 842207 846841 846842 846844 846845 846847 846848 846849 846850 846851 846853 846855 942926 942943 942944 942945 942946 942947 942950 942951 942953 942954 942960 942964 943554 1044584 1044800 1045871 1046151 1046152 1046159 1046162 1046165 1140204 1140205 1140659 1140660 1140662 1142190 1142191 1143092 1143093 1143436 1143437 1144693 1144694 1144696 1145106 1145400 1145478 1146161 1146982 1240188 1240189 1241117 1241949 1242400 1243129 1243584 1243585 1244442 1244882 1246186 1246187 1340352 1340493 1340494 1341007 1341164 1341880 1342685 1342858 1344252 1344707 1440353 1442562 1443626 544043 645847

Other Recalls from Stryker Endoscopy

Recall # Classification Product Date
Z-1566-2016 Class II PneumoSure XL High Flow Insufflator and Pneumo... Mar 28, 2016
Z-1831-2015 Class II SERFAS 90 degree Energy Probe, Part Number 279-... Jun 3, 2015
Z-0631-2015 Class II PKG, PENNINGTON FORCEPS, P/N 0250080242. Lapa... Nov 17, 2014
Z-0646-2015 Class II PKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250... Nov 17, 2014
Z-0679-2015 Class II PKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparo... Nov 17, 2014

Frequently Asked Questions

Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.