Philips IP5 Information Portal The Expression Information Portal (Model IP5), a display and de...
FDA Device Recall #Z-2270-2014 — Class II — June 30, 2014
Recall Summary
| Recall Number | Z-2270-2014 |
| Classification | Class II — Moderate risk |
| Date Initiated | June 30, 2014 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Invivo Corporation |
| Location | Orlando, FL |
| Product Type | Devices |
| Quantity | 573 devices. |
Product Description
Philips IP5 Information Portal The Expression Information Portal (Model IP5), a display and device controller for Philips MRI patient monitoring systems, is intended for use by trained health care professionals to remotely monitor the vital signs of patients undergoing MRI procedures.
Reason for Recall
The wireless function of the IP5 radio could be interrupted causing loss of wireless communication between the IP5 and host monitoring system.
Distribution Pattern
Worldwide distribution: US (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, MN, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, Hawaii, Alaska, and District of Columbia; and countries of: Australia, Austria, Denmark, France, Germany, Ireland, Netherlands, Norway, Poland, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
Lot / Code Information
Product: 865471 Number 453564251511 Serial Numbers: US23200121, US23200122, US23200123, US23200124, US23200125, US23200126, US23200127, US23200128, US23200129, US23200130, US23200131, US23200132, US23200133, US23200134, US23200135, US23200136, US23200137, US23200138, US23200139, US23200140, US23200141, US23200142, US23200143, US23200144, US23200145, US23200146, US23200147, US23200148, US23200149, US23200150, US23200151, US23200152, US23200153, US23200154, US23200155, US23200156, US23200157, US23200158, US23200159, US23200160, US23200161, US23200162, US23200163, US23200164, US23200165, US23200166, US23200167, US23200168, US23200169, US23200170, US23200172, US23200173, US23200176, US23200177, US23200178, US23200179, US23200180, US23200181, US23200182, US23200183, US23200184, US23200188, US23200189, US23200190, US23200191, US23200192, US23200193, US23200194, US23200195, US23200196, US23200197, US23200198, US23200199, US23200200, US23200201, US23200202, US23200203, US23200204, US23200205, US23200206, US23200207, US23200208, US23200209, US23200210, US23200211, US23200212, US23200213, US23200214, US23200215, US23200216, US23200217, US23200218, US23200219, US23200220, US23200221, US23200222, US23200223, US23200224, US23200225, US23200226, US23200227, US23200228, US23200229, US23200230, US23200231, US23200232, US23200233, US23200234, US23200235, US23200236, US23200237, US23200238, US23200239, US23200240, US23200241, US23200242, US23200243, US23200244, US23200245, US23200246, US23200247, US23200248, US23200249, US23200250, US23200251, US23200252, US23200253, US23200254, US23200255, US23200256, US23200257, US23200258, US23200259, US23200260, US23200261, US23200262, US23200263, US23200264, US23200265, US23200266, US23200267, US23200268, US23200269, US23200270, US23200271, US23200272, US23200273, US23200274, US23200275, US23200276, US23200277, US23200278, US23200279, US23200280, US23200281, US23200282, US23200283, US23200284, US23200285, US23200286, US23200287, US23200288, US23200289, US23200290, US23200291, US23200292, US23200293, US23200294, US23200295, US23200296, US23200297, US23200298, US23200299, US23200300, US23200301, US23200302, US23200303, US23200304, US23200305, US23200306, US23200307, US23200308, US23200309, US23200310, US23200311, US23200312, US23200313, US23200314, US23200315, US23200316, US23200317, US23200318, US23200319, US23200320, US23200321, US23200322, US23200323, US23200324, US23200325, US23200326, US23200327, US23200328, US23200329, US23200330, US23200331, US23200332, US23200333, US23200334, US23200335, US23200336, US23200337, US23200338, US23200339, US23200340, US23200341, US23200342, US23200343, US23200344, US23200345, US23200346, US23200347, US23200349, US23200350, US23200351, US23200352, US23200353, US23200354, US23200356, US23200357, US23200358, US23200359, US23200360, US23200361, US23200362, US23200363, US23200364, US23200365, US23200366, US23200367, US23200368, US23200369, US23200370, US23200371, US23200372, US23200373, US23200374, US23200375, US23200376, US23200377, US23200378, US23200379, US23200380, US23200381, US23200382, US23200383, US23200385, US23200386, US23200387, US23200388, US23200389, US23200390, US23200391, US23200393, US23200394, US23200395, US23200397, US23200398, US23200399, US23200400, US23200401, US23200402, US23200403, US23200405, US23200406, US23200407, US23200408, US23200409, US23200410, US23200411, US23200413, US23200415, US23200416, US23200419, US23200421, US23200422, US23200423, US23200424, US23200425, US23200427, US23200428, US23200429, US23200431, US23200434, US34600510, US34600511, US34600512, US34600513, US34600514, US34600515, US34600516, US34600517, US34600518, US34600519, US34600520, US34600521, US34600522, US34600523, US34600524, US34600525, US34600526, US34600528, US34600529, US34600530, US34600531, US34600532, US34600533, US34600534, US34600535, US34600536, US34600537, US34600538, US34600539, US34600540, US34600541, US34600542, US34600543, US34600544, US34600545, US34600546, US34600547, US34600548, US34600549, US34600550, US34600551, US34600552, US34600553, US34600554, US34600555, US34600556, US34600557, US34600558, US34600559, US34600560, US34600561, US34600562, US34600563, US34600564, US34600565, US34600566, US34600567, US34600568, US34600569, US34600570, US34600571, US34600572, US34600573, US34600574, US34600575, US34600576, US34600577, US34600578, US34600579, US34600580, US34600581, US34600582, US34600583, US34600584, US34600585, US34600586, US34600587, US34600588, US34600589, US34600590, US34600591, US34600592, US34600592, US34600593, US34600594, US34600595, US34600596, US34600597, US34600598, US34600599, US34600600, US34600601, US34600602, US34600603, US34600604, US34600605, US34600606, US34600607, US34600608, US34600609, US34600610, US34600611, US34600612, US34600613, US34600614, US34600615, US34600616, US34600617, US34600618, US34600619, US34600620, US34600621, US34600622, US34600623, US34600624, US34600625, US34600626, US34600627, US34600628, US34600629, US34600630, US34600631, US34600632, US34600633, US34600634, US34600635, US34600636, US34600637, US34600638, US34600639, US34600640, US34600641, US34600642, US34600643, US34600644, US34600645, US34600646, US34600647, US34600648, US34600649, US34600650, US34600651, US34600652, US34600653, US34600654, US34600655, US34600656, US34600657, US34600658, US34600659, US34600660, US34600661, US34600662, US34600663, US34600664, US34600665, US34600666, US34600667, US34600668, US34600669, US34600670, US34600671, US34600672, US34600673, US34600674, US34600675, US34600676, US34600677, US34600678, US34600679, US34600680, US34600681, US34600682, US34600683, US34600684, US34600685, US34600686, US34600687, US34600688, US34600689, US34600690, US34600691, US34600692, US34600693, US34600694, US34600695, US34600696, US34600697, US34600698, US34600699, US34600700, US34600701, US34600702, US34600703, US34600704, US34600705, US34600706, US34600707, US34600708, US34600709, US34600710, US34600711, US34600712, US34600713, US34600714, US34600715, US34600716, US34600717, US34600718, US34600719, US34600720, US34600721, US34600722, US34600723, US34600724, US34600729, US34600730, US34600731, US34600732, US34600733, US34600734, US34600735, US34600736, US34600737, US34600738, US34600739, US34600740, US34600741, US34600742, US34600743, US34600744, US34600745, US34600747, US34600748, US34600749, US34600750, US34600751, US34600752, US34600753, US34600754, US34600755, US34600756, US34600757, US34600758, US34600759, US34600760, US34600761, US34600762, US34600763, US34600764, US34600765, US34600766, US34600767, US34600768, US34600769, US34600770, US34600771, US34600772, US34600773, US34600774, US34600775, US34600776, US34600777, US34600778, US34600779, US34600780, US34600781, US34600782, US34600783, US34600784, US34600786, US34600787, US34600788, US34600789, US34600790, US34600791, US34600792, US34600794, US34600795, and US34600796
Other Recalls from Invivo Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0126-2021 | Class III | MR Coils The MR Coil is intended to be used ... | Sep 14, 2020 |
| Z-1440-2020 | Class II | Monitor for the Functional Imaging System for M... | Feb 6, 2020 |
| Z-0462-2019 | Class II | Xper Flex Cardio Physiomonitoring System, Softw... | Sep 24, 2018 |
| Z-3192-2018 | Class II | Xper Flex Cardio Physiomonitoring System Model ... | Aug 3, 2018 |
| Z-1867-2018 | Class II | PHILIPS Xper Flex Cardio Physiomonitoring Syste... | Mar 14, 2018 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.