MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem D...
FDA Device Recall #Z-1528-2015 — Class II — August 8, 2014
Recall Summary
| Recall Number | Z-1528-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | August 8, 2014 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Beckman Coulter, Inc. |
| Location | West Sacramento, CA |
| Product Type | Devices |
| Quantity | Approx ~(216240 boxes U.S. and 84249 boxes OUS) total |
Product Description
MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg Combo 50, Catalog B1017-406 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
Reason for Recall
Technical Support Bulletin issued in February 2013 did not mention performance limitation for Proteus mirabilis and Imp when using the CLSIM100-S20 revised interpretive criteria.
Distribution Pattern
Worldwide Distribution: US (Nationwide) and to countries of: Argentina, Bahamas, Brazil, Canada, China, Ecuador, Egypt, Germany, Lithuania, Mexico, Paraguay, Peru Poland, Portugal, Romania, Russian Federation, Singapore, South Africa, Spain, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates, Uruguay, and Venezuela.
Lot / Code Information
Catalog number: B1017-406; Siemens Material Number (SMN): 10444700,
Other Recalls from Beckman Coulter, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1235-2026 | Class II | CellMek SPS Sample Preparation System, REF: C44... | Dec 23, 2025 |
| Z-1025-2026 | Class II | Beckman Coulter UniCel Dxl 600, Part numbers: ... | Nov 26, 2025 |
| Z-1026-2026 | Class II | Beckman Coulter UniCel Dxl 800, Part numbers: ... | Nov 26, 2025 |
| Z-1221-2026 | Class II | COULTER DxH Diluent, REF 628017 COULTER DxH EC... | Nov 10, 2025 |
| Z-0889-2026 | Class II | Access 2 Reaction Vessels, individual, disposab... | Nov 7, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.