Brilliance CT Big Bore Oncology Whole Body Computed Tomography X-Ray System & Brilliance CT Big ...
FDA Recall #Z-1745-2014 — Class II — December 19, 2013
Product Description
Brilliance CT Big Bore Oncology Whole Body Computed Tomography X-Ray System & Brilliance CT Big Bore Radiology Whole Body Computed Tomography X-Ray System. Product Usage: The Brilliance CT Big Bore is a Whole Body Computed Tomography X-Ray System which produces cross-sectional body images by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason for Recall
During internal testing, the Brilliance CT Big Bore was found to be out of tolerance for radio frequency emissions. Specifically, at the 48MHZ frequency, the testing indicated the Brilliance CT Big Bore was 3.5dB uV/meter higher than the applicable IEC 60601-1-2 standard specification.
Recalling Firm
Philips Medical Systems (Cleveland) Inc — Cleveland, OH
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
454 Units
Distribution
Worldwide Distribution - US (nationwide) in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI & WV. and countries of Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Ecuador, France, French Polynesia, Gabon, Germany, India, Indonesia, Israel, Italy, Jordan, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Zealand, Oman, Palestine, Philippines, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syrian Arab Republic, Thailand, Ukraine & United Kingdom.
Code Information
Brilliance CT Big Bore Oncology: Model #728243; Serial #s: 7185, 7187-7189, 7191-7199, 7201,-7243, 7245-7251, 7253-7267, 7269, 7270, 7272-7297, 7299, 7300, 7302, 7304-7317, 7319-7374, 7376-7379, 7381, 7382, 7385, 7387-7389, 7391-7393, 7395, 7397-7401, 7403-7406, 7409-7429, 7431-7433, 7435-7438, 7440-7451, 7454-7457, 7459, 7462-7485, 7487-7509, 7511-7513, 7515-7519, 7521-7524, 7526-7528, 7530-7532, 7534-7537, 7539-7542, 7544, 7545, 7547, 7549-7556, 7558-7562, 7564-7566, 7568, 7569, 7573, 7574, 7577-7582, 7584, 7585, 7588-7593, 7596-7599, 7601-7603, 7605, 7608, 7609, 7611-7613, 7615-7618, 7620, 7622-7624, 7626-7630, 7634, 7635, 7637-7648 & 75000. Brilliance CT Big Bore Radiology: Model #728244; Serial #s: 7190, 7200, 7244, 7252, 7268, 7271, 7301, 7318, 7375, 7380, 7383, 7384, 7386, 7390, 7396, 7402, 7407, 7408, 7427, 7430, 7439, 7452, 7453, 7458, 7486, 7510, 7514, 7520, 7525, 7529, 7533, 7538, 7543, 7546, 7557, 7570-7572, 7575, 7576, 7583, 7586, 7587, 7595, 7600, 7604, 7607, 7610, 7614, 7619, 7621, 7625, 7631-7633 & 7636.
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated