ComboMap Intravascular Flow and Pressure System; Product/Part number: Combo Map - 802745-001. ...

FDA Device Recall #Z-2045-2015 — Class II — September 23, 2013

Recall Summary

Recall Number Z-2045-2015
Classification Class II — Moderate risk
Date Initiated September 23, 2013
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Volcano Corp
Location San Diego, CA
Product Type Devices
Quantity 227 total units - all models

Product Description

ComboMap Intravascular Flow and Pressure System; Product/Part number: Combo Map - 802745-001. Intravascular Imaging and Pressure System for evaluation of blood flow, perfusion and vascular structures.

Reason for Recall

A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.

Distribution Pattern

Worldwide Distribution - US Nationwide and the countries of Belgium, United Kingdom, Denmark, Netherlands, Sweden, Iceland, France, Saudi Arabia, Australia, Philippines and Thailand.

Lot / Code Information

Product/Part numbers: Combo Map - 802745-001. All software versions

Other Recalls from Volcano Corp

Recall # Classification Product Date
Z-2221-2025 Class II Volcano Visions Digital IVUS Catheter: PV.014P ... Jun 13, 2025
Z-1270-2022 Class II Philips OmniWire Pressure guide wire REF 89185J... May 16, 2022
Z-1301-2022 Class III Volcano Visions PV. 035 Digital IVUS Catheter R... May 13, 2022
Z-1302-2022 Class III Volcano Visions PV. 035 Digital IVUS Catheter R... May 13, 2022
Z-0917-2022 Class III MM-TSM (Touch Screen Module/Secondary Controlle... Feb 3, 2022

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.