Hematoxylin II, Ventana Part Number 790-22087, Roche GMMI 05277965001 in vitro diagnostic
FDA Recall #Z-0738-2018 — Class I — December 19, 2017
Product Description
Hematoxylin II, Ventana Part Number 790-22087, Roche GMMI 05277965001 in vitro diagnostic
Reason for Recall
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
Recalling Firm
Ventana Medical Systems Inc — Oro Valley, AZ
Classification
Class I — Reasonable probability of serious adverse health consequences or death.
Product Type
Devices
Product Quantity
71640 units
Distribution
Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,
Code Information
UDI 04015630972135, Lot Numbers: Y10759, Y17402, Y21312, Y13938, Y17403, Y22561
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated