GE Healthcare Precision RXi 23A/32A Analog System
FDA Device Recall #Z-2121-2015 — Class II — June 12, 2015
Recall Summary
| Recall Number | Z-2121-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | June 12, 2015 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | GE Healthcare |
| Location | Waukesha, WI |
| Product Type | Devices |
| Quantity | 788 (701 US; 87 OUS) |
Product Description
GE Healthcare Precision RXi 23A/32A Analog System
Reason for Recall
A reported incident of a monitor that fell from the suspension. A fall of a monitor could result in an injury to a patient or operator. Recall affects the single and dual LCD monitor suspensions used on Precision 500D Classical R & F system, Precision RXi 23A/32A Analog System, Precision RXi Digital system and Precision MPi fluoroscopic imaging systems.
Distribution Pattern
Worldwide Distribution - US Nationwide Distribution including GU and DC., and to the countries of : Australia, Bulgaria, Canada, Chile, Dominican Republic, Germany, Ireland, Italy, Kuwait, Lebanon, Portugal, Romania, Russia, Saudi Arabia, Slovakia, Spain, Taiwan,Turkey, United Kingdom, Venezuela and Yemen.
Lot / Code Information
Mfg Lot or Serial # 00000001386MC2 00000002109MC7 00000001010MC8 00000001337MC5 00000002212MC9 00000001720MC2 00000001098MC3 00000001454MC8 00000001082MC7 00000001266MC6 00000001656MC8 00000003661MC6
Other Recalls from GE Healthcare
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1475-2026 | Class II | GE Healthcare Centricity Universal Viewer Zero ... | Jan 16, 2026 |
| Z-1487-2025 | Class II | GE Healthcare Centricity Centricity PACS-IW wit... | Feb 19, 2025 |
| Z-1484-2025 | Class II | GE Healthcare Centricity PACS-IW (PACS-IW), Mod... | Feb 19, 2025 |
| Z-1486-2025 | Class II | GE Healthcare Centricity Cardiology CA1000 (CA1... | Feb 19, 2025 |
| Z-1485-2025 | Class II | GE Healthcare Centricity Radiology RA600 (RA600... | Feb 19, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.